Output 1, to select the conguration of the digital output of transistor 1.
Output 2, to select the conguration of the digital output of transistor 2.
Back, returns to the home screen of the setup menu Figure 113.
Main Menu, back to the main menu, Figure 25.
Use the keys
and to select the different options.
To conrm the selection, press the key
.
If no key is pressed for 5 minutes, the display screen changes automatically to the default
screen, which displays the voltage measurement of 4 parameters.
After selecting the output to congure, you can congure the transistor output mode, Alarm or
Impulse output, and the parameters for each mode.
Alarm, to select the alarm mode of the transistor digital output.
Impulses, to select the impulse mode of the transistor digital output.
5.6.16.1.-
Alarm mode of transistor digital output
When selecting the alarm mode,
Figure 131, the conguration parameters are:
The variable code that controls the transistor.
The pre-alarm value.
The minimum value below which the transistor is activated.
The maximum value above which the transistor is activated.
The delay in the connection and disconnection of the transistor.
The status of the outputs.
The latch.
Figure 131: Conguration screen of transistor digital outputs (alarm mode).