Instruction manual
Getest
Specifications
Mains supply
Voltage 230V+/-20% 50-60Hz
Current 15A AC
Note: Required low impedance in mains supply line for getting maximum performance, as up to
320A 20ms current surges may appear when requesting maximum currents. Mains supply wire
section recommended 2.5mm
2
or bigger, in order to have a low line resistance.
Nota2
:
We recommend to protect the mains power line with a circuit breaker : 50A curve B, or
30A curve C, or 16A curve D, as per EN60898
Current injection
Max output voltage: 600VAC
Max output current: 50 A.
Max power: (30KVA).
Minimum value limits:
I
max
= (30000/R
load
) or I
max
= 600/R
load
Voltage meter
Type: true RMS
Input impedance: 10M or 1K, selectable
Resolution:
700VAC scale – 0.5VAC resolution
280VAC scale – 0.1VAC resolution
105VAC scale – 0.1AC resolution
70VAC scale – 0.1VAC resolution
30VAC scale – 0.1VAC resolution
0.8VAC scale – 0.01VAC resolution
Max voltage: 700VAC
Scales: Auto (0.8, 30, 70, 105, 280, 700)
Accuracy : 1% reading + 2 digits
Current meter
Type : true RMS
Input impedance: less than 0.1
Resolution:
100AAC scale – 0.1AAC resolution
10AAC scale – 0.01AAC resolution
Max current: 100AAC
Scales: Auto (10, 100)
Accuracy : 2% reading + 2 digits
Resistance meter
Range : 0 - 150
Accuracy : 2.5% reading + 2 digits
Communications
UART protocol, key protected, over Bluetooth v1.2
Class I
Tx Power: +18dBm
Sensitivity: -88dBm
Regulatory approvals:
FCC Part 15 Subpart C Section 15.247
ETSI EN 300 328
ETSI EN 301 489-1
Seitenansicht 16