59
CVM
k
2
CONFIGURATION
FEATURES VALUE UNIT
ANALOGUE OUTPUTS
Maximum internal voltage 12 VDC
Maximum external voltage 24 VDC
Rated output range 0/4 ... 20 mA d.c
Linearity 1 %
Load resistance range < 300 Ω
Load resistance range (external supply) < 600 Ω
Resolution 4.000 points
ANALOGUE INPUTS
Type of measurement Current
Rated input range 0/4 ... 20 mA mA d.c
Measurement precision 1 %
Input impedance 200 Ω
CONNECTIONS
Wire cross section (Cu) 0,05..1 (AWG 30...18) mm
2
Maximum torque 0,3 Nm
MECHANICAL
Terminals protection IP 20
4.7.3.5. Features
4.7.3.4. Expansion card parameters
To see the parameters of the expansion card, you have to intro in menu, select cards, and go
to the card to see the parameters.
The picture shows the status of the inputs of the cards or the number of impulses that has
congured each one.